TOXIC GAS METERS BASIC INFORMATION AND TUTORIALS


What is a toxic gas meter?


Description and Application.
This analyzer uses an electrochemical voltametric sensor or polarographic cell to provide continuous analyses and electronic recording. In operation, sample gas is drawn through the sensor and absorbed on an electrocatalytic sensing electrode, after passing through a diffusion medium.

An electrochemical reaction generates an electric current directly proportional to the gas concentration. The sample concentration is displayed directly in parts per million. Since the method of analysis is not absolute, prior calibration against a known standard is required.

Exhaustive tests have shown the method to be linear; thus, calibration at a single concentration, along with checking the zero point, is sufficient.

Types: Sulfur dioxide, hydrogen cyanide, hydrogen chloride, hydrazine, carbon monoxide, hydrogen sulfide, nitrogen oxides, chlorine, and ethylene oxide. These can be combined with combustible gas and oxygen meters.

Calibration. 
Calibrate the direct-reading gas monitor before and after each use in accordance with the manufacturers instructions and with the appropriate calibration gases.

Special Considerations.

• Interference from other gases can be a problem. See manufacturers literature.

• When calibrating under external pressure, the pump must be disconnected from the sensor to avoid sensor damage. If the span gas is directly fed into the instrument from a regulated pressurized cylinder, the flow rate should be set to match the normal sampling rate.

• Due to the high reaction rate of the gas in the sensor, substantially lower flow rates result in lower readings. This high reaction rate makes rapid fall time possible simply by shutting off the pump. Calibration from a sample bag connected to the instrument is the preferred method.

No comments:

Post a Comment